University of Washington

Department of Rehabilitation Medicine

http://rehab.washington.edu/research/articles/showref.asp?id=4583


Search Again

Title:

StepWatch stride counting: accuracy, precision, and prediction of energy expenditure in children

Author(s):

Bjornson, KF, Yung, D, Jacques, K, Burr, RL, Christakis, D

Year:

2012

Publication Info:

Journal of Pediatric Rehabilitation Medicine, 5:7-14

Abstract:

INTRODUCTION: Accuracy and precision of the StepWatch (SW) was tested in typically developing youth (TDY) with the prediction of oxygen consumption derived from stride counts.
METHODS: Ten TDY (5 boys, 5 girls) with an average age 14.1 years (SD=2.2) enrolled. Participants underwent treadmill cardiopulmonary exercise testing (CPET) wearing two SW devices at 1, 2, 3, and 4 miles per hour (mph).
RESULTS: Average sign-corrected disagreement for strides counted between the 2 monitors was 0.4 (0.52), 0.2 (0.42), 0.1 (0.32), and 0.1 (0.32) for 1, 2, 3, and 4 mph, respectively. The ratio of SW counts to observed manual counts averaged 100.03% with ICC=0.995. Oxygen consumption equations were derived from resting and walking VO_{2}, age, gender, and stride counts.
CONCLUSIONS: Excellent accuracy and precision was documented for treadmill walking speeds up to 4 mph while VO_{2} per number of strides taken appears to decrease with age and is lower in girls.

Link to Article:

http://www.ncbi.nlm.nih.gov/pubmed/?term=10.3233%2FPRM-2011-0186